SPMwikidI/dV spectrum by Createc system 2017-04-06
How to obtain dI/dV mapping singal by Createc software
1:Set the interal lock-in parameters
For point spectrum:RC(Hz): should be smaller, such as 0.7Hz. Lockin-Mode:Internal+scan off (also works for Internal mode)
For mapping:
RC(Hz): should be larger, such as 14Hz. Lockin-Mode:Internal
All the parameters include Frequency(Hz) and Amplitude you should play around according to yous system conditions
Then press Autophase button and you will have a Phase_1 value
2:Set Vertical Manipulation parameters( For point spectrum)
A typical Au(111) dI/dV point spectrum setting(LN2 temperature) was showed in the next picture*You need to play around the parameters according to your system conditions
Channels:you should select the dI/dV channels to collect dI/dV singals for interal lock-in
VertChannelForm:Click this button to show the X/Y graph for your singal
V-Z Table:You should set the bias range here
3:Scannig conditons for dI/dV mapping
For dI/dV mapping(By internal lockin),you should select the dI/dV channel in Parameter>scan panel.The scan speed should be slower, typical 1s/lines.(also should be adjusted by yourself)
After you have found the correct lockin parameters in step 1. You can start STM scanning with dI/dV singal.
Press Ctrl+N to open the window which will show the dI/dV mapping channel.